Apostilb: Luminance Measurement in Vision Science

The apostilb is a non-SI unit of luminance equal to 1/π candela per square meter. Explore its definition, history, mathematical foundations, and clinical use in perimetry.

Dr. Mohammed Looti An academic and researcher specialized in
PUBLISHED
Scientifically Reviewed · Dr. Marwa Abd-Alazim · November 6, 2026
Medically & Scientifically Reviewed Verified: November 6, 2026
Dr. Marwa Abd-Alazim PhD in Psychology
Faculty Member • University of Kerbala
Review Criteria & Clinical Standards

This content undergoes rigorous scientific peer-review and medical editorial standards at Arab Psychology Network to ensure clinical accuracy, validity, and compliance with evidence-based guidelines from leading psychological and healthcare authorities (APA / WHO).

In visual psychophysics, ophthalmic diagnostics, and optical engineering, the precise quantification of visible light emitted or reflected from physical surfaces is foundational to understanding sensory perception. The apostilb represents a specialized, historical non-SI unit of luminance that bridged theoretical radiometry and physiological optics throughout the twentieth century. While modern scientific conventions have largely standardized around the International System of Units (SI) measure of candela per square meter, the apostilb remains deeply embedded in the historical literature of sensory physiology and continues to serve as an indispensable metric in clinical visual field testing and perimeter calibration worldwide.

Apostilb

1. Concise Definition

The apostilb (symbol: asb), occasionally designated as the blondel, is a non-SI unit of luminance defined as the surface luminance of an ideal, perfectly diffusing (Lambertian) surface emitting or reflecting a luminous flux of one lumen per square meter. In mathematical equivalence within the International System of Units, one apostilb is precisely equal to 1/π candela per square meter (approximately 0.3183 cd/m² or nit), or 10−4 lamberts.

Unlike absolute directional luminance measures that define luminous intensity emitted per unit of projected surface area into a given solid angle, the apostilb was devised specifically to simplify practical calculations involving uniformly diffusing surfaces. By incorporating the geometric hemispherical integration constant π directly into the unit's definition, the apostilb equates the numerical value of luminous exitance (measured in lumens per square meter or lux) directly to surface luminance. Consequently, an ideal diffuse reflector possessing an illuminance of one lux exhibits a luminance of exactly one apostilb under unity reflectance conditions.

Although formally depreciated by international metrological organizations such as the International Commission on Illumination (Commission Internationale de l'Éclairage, CIE) and the International Bureau of Weights and Measures (BIPM), the apostilb maintains an enduring clinical presence. Most prominently, it defines the target stimulus and background luminance standards utilized in standardized visual field perimetry, such as Goldmann kinetic perimetry and automated static perimetry protocols.

2. Etymology & Linguistic Origin

The term apostilb is a compound neologism derived from classical Greek roots, constructed within the late nineteenth and early twentieth-century traditions of international scientific nomenclature. The prefix apo- (απó) denotes “away from,” “off,” or “derived from,” signaling a secondary or modified derivation. The root stilb is taken directly from the ancient Greek verb stilbein (στíλβειν), which translates as “to glitter,” “to glisten,” or “to shine brightly.” The stilb had previously been established in physical photometry as the unit representing one candela per square centimeter.

Because the stilb represented an inconveniently large magnitude of luminance for everyday human environments, visual ergonomics, and architectural illumination, European photometrists established the derivative prefix apo- to denote a substantially scaled-down, diffuse counterpart. By dividing the stilb by both a spatial factor of 10,000 (converting square centimeters to square meters) and the geometric diffuse dispersion factor π, the apostilb represented a fraction (10−4/π) of a stilb, far better aligned with natural, comfortable human visual environments.

In French scientific literature and select international conventions during the mid-twentieth century, the unit was alternately designated as the blondel in honor of the pioneering French physicist and engineer André-Eugène Blondel (1863–1938), whose profound mathematical contributions systematized modern physical photometry and optical terminology. Over time, the designation apostilb became the dominant term within central European, Scandinavian, and ophthalmic scientific discourse.

3. Pronunciation & Grammatical Form

In standardized scientific English, apostilb is pronounced phonetically as /æp.é™.stæªlb/ or /é™ËpË’stæªlb/. The primary lexical stress is traditionally placed upon the second syllable (“-pos-”) in British English and frequent academic usage, though variant pronunciation stressing the first syllable (“ap-”) is documented within clinical North American environments.

Grammatically, apostilb functions exclusively as an inanimate countable noun. Its standard plural form is apostilbs (e.g., “a background illumination set at 31.5 apostilbs”). The standardized scientific symbol is asb, which is written without an apostrophe, plural suffix, or terminal period, separated by a non-breaking space from the preceding numerical quantity (for example, “10 asb” or “10,000 asb”). In older continental European literature, occasional abbreviations such as as or asbl appear, though these are non-standard and strongly discouraged in contemporary historical citations.

4. Detailed Conceptual Explanation

To fully grasp the theoretical concept of the apostilb, one must examine the critical physical distinction between luminous flux, illuminance, luminous exitance, and luminance. Luminous flux, quantified in lumens, describes the total quantity of visible light energy emitted by a source per second, weighted according to the spectral sensitivity of the human visual system (the photopic luminous efficiency function). When luminous flux falls onto an external surface, the areal density of that incident light is known as illuminance, quantified in lux (lumens per square meter). Conversely, when that surface reflects or emits that light outward into space, the total flux leaving per unit area regardless of direction is luminous exitance, also measured in lumens per square meter.

However, the human visual system does not perceive luminous exitance directly; rather, it responds to luminance. Luminance describes the luminous intensity (flux per unit solid angle, measured in candelas or lumens per steradian) emerging from a surface in a given direction, divided by the projected area of the surface as viewed from that perspective. The projected area accounts for the visual foreshortening described by the cosine of the viewing angle relative to the surface normal. Thus, luminance governs how bright an extended surface appears to an observer, irrespective of the observer's viewing distance.

For an ideal matte surface that obeys Lambert's cosine law, light is scattered with uniform apparent brightness across all viewing angles. While the luminous intensity drops off with the cosine of the angle away from the normal, the projected area decreases by the exact same cosine factor, rendering luminance constant across the entire viewing hemisphere. When one mathematically integrates this cosine-weighted distribution across the complete hemisphere (a solid angle of 2π steradians), the integral equates precisely to π steradians. Therefore, the relationship between luminous exitance (M) and luminance (L) for a Lambertian surface is expressed by the fundamental photometric equation:

M = π × L

In standard SI units, if a Lambertian surface has a luminance L of 1 candela per square meter, its total luminous exitance M is π lumens per square meter. Recognizing that practitioners in architecture, lighting design, and clinical medicine frequently found multiplying or dividing by irrational values of π cumbersome in practical field work, the apostilb was created. By defining 1 apostilb as (1/π) cd/m², the factor of π cancels out completely:

M (in lumens/m²) = L (in apostilbs) × ρ

where ρ represents the diffuse reflectance coefficient of the material (ranging from 0.0 for a perfect absorber to 1.0 for an ideal diffuse white reflector). Thus, an architect or clinician shining 100 lux of light upon a perfectly white diffuse card (ρ = 1.0) could instantly state, without any trigonometric calculation, that the card exhibited a luminance of exactly 100 apostilbs.

5. Historical Development

The origins of the apostilb trace back to the rapid evolution of photometric science between the late nineteenth century and the early decades of the twentieth century. Following the seminal theoretical foundations established by Pierre Bouguer and Johann Heinrich Lambert in the eighteenth century, practical illumination engineering expanded dramatically with the rise of widespread gas and incandescent electric lighting. During this era, national scientific bodies developed disparate, non-standardized systems of units based on varied luminous intensity standards, including the British standard sperm candle, the German Hefner lamp (Hefnerkerze), and the French Carcel lamp.

In the 1920s and 1930s, the Commission Internationale de l'Éclairage worked extensively to unify optical and photometric terminology across borders. Central European engineers, particularly within the German standards organization (Deutscher Normenausschuss, now DIN), favored a metric system that streamlined calculations involving diffuse reflection and interior lighting conditions. The apostilb emerged during this interwar period as a standard unit of luminance across Germany, Austria, Switzerland, and neighboring European countries, formalized under industrial lighting standards such as DIN 5032 and DIN 5035.

A transformative milestone in the history of the apostilb occurred in the late 1940s through the collaborative work of Swiss ophthalmologist Hans Goldmann and instrument manufacturer Haag-Streit. Goldmann recognized that accurate, reproducible mapping of the human visual field required absolute standardization of stimulus contrast and background adaptation. When designing the Goldmann bowl perimeter, Goldmann calibrated the hemispherical white bowl to an adaptation luminance of 31.5 apostilbs (which corresponds to exactly 10 candelas per square meter) and established standardized stimulus luminance gradations calibrated across logarithmic steps up to 1,000 and 10,000 apostilbs.

During the subsequent transition of global metrology to the Système International (SI) following the 11th General Conference on Weights and Measures (CGPM) in 1960, non-SI photometric units—including the stilb, apostilb, lambert, and foot-lambert—were systematically slated for deprecation in favor of the coherent unit candela per square meter (cd/m²). Despite formal obsolescence in engineering standards by the late 1970s, the apostilb remained permanently entrenched within clinical ophthalmology and optometry due to decades of accumulated visual field norm databases and clinical trial data.

6. Theoretical Foundations

The theoretical framework sustaining the apostilb resides at the intersection of classical radiometry, biological vision science, and Lambertian surface physics. Photometry differs fundamentally from radiometry because it weights physical radiant flux across each wavelength according to the human eye's sensitivity, modeled by the standard photopic spectral luminous efficiency function, V(λ). The luminous flux Φv is calculated via:

Φv = Km ∫ Φe,λ(λ) V(λ) dλ

where Km is the maximum spectral luminous efficacy (approximately 683 lumens per watt at 555 nanometers in photopic conditions), and Φe,λ(λ) is the spectral radiant flux. The apostilb operates entirely within this human-referenced visual framework.

Visual perception of luminance depends upon retinal illuminance, traditionally measured in trolands (Td), defined as the product of visual field luminance (in cd/m²) and pupil area (in mm²). Because human spatial contrast sensitivity adheres closely to the Weber–Fechner law over a vast photopic operating range, the visual system detects objects based on relative luminance contrast (ΔL / L) rather than absolute radiant power. In this context, Weber's fraction dictates that:

ΔL / L = k

where ΔL is the difference in luminance between a target stimulus and its surrounding background, L is the background adaptation luminance, and k is a constant. By calibrating the visual background L to a stable level of 31.5 asb (10 cd/m²), the visual apparatus is held securely in the photopic adaptation state where Weber's law holds constant, ensuring that sensory thresholds remain stable across clinical assessments.

Furthermore, theoretical models of diffuse scattering explain why the apostilb was mathematically intuitive. Real-world surfaces, such as diffuse painted walls or the matte white bowl of an optical perimeter, act as near-ideal Lambertian scatterers through multiple internal subsurface reflections. Because the reflected radiance is completely isotropic, the luminance L across any emission angle θ is invariant. The mathematical integration over the entire solid angle hemisphere yields the definitive constant π, validating the apostilb's foundational identity as an exitance-equivalent luminance unit.

7. Key Components, Types & Dimensions

Understanding the apostilb requires analyzing its dimensional structure, its physical conversions, and its relationships with other legacy and modern photometric units:

  • Dimensional Formula: In SI base dimensions, luminance possesses the dimensional formula J · L−2, where J represents luminous intensity (candela) and L represents length (meter). The apostilb shares this dimensional identity, scaled by the dimensionless mathematical factor 1/π.
  • Conversion to SI Unit (Candela per Square Meter):

    1 asb = 1/π cd/m² ≈ 0.318309886 cd/m² (nit).

    Conversely, 1 cd/m² = π asb ≈ 3.14159265 asb.
  • Conversion to the Stilb (sb):

    The stilb represents 1 cd/cm² (10,000 cd/m²). Therefore, 1 asb = 10−4/π sb ≈ 3.183 × 10−5 sb.
  • Conversion to the Lambert (L):

    The lambert is defined as (1/π) cd/cm². Consequently, 1 asb = 10−4 L = 0.0001 L.
  • Conversion to the Foot-Lambert (fL):

    The foot-lambert is the imperial diffuse counterpart to the apostilb, representing (1/π) candela per square foot. Because one square foot equals approximately 0.092903 square meters, 1 fL ≈ 10.7639 asb, and 1 asb ≈ 0.092903 fL.
  • Scotopic Variants (Skot and Bril):

    In specialized dark-adaptation and military night-vision research, German scientists introduced the skot (symbol: sk) as an equivalent diffuse unit for scotopic vision, defined as 10−3 apostilb under scotopic spectral luminous efficiency V'(λ). The bril represented a further sub-threshold scotopic division (10−7 asb).

8. Examples & Illustrative Cases

To contextualize the physical magnitude of an apostilb, consider standard environments encountered in clinical diagnostics, architectural environments, and natural outdoor illumination.

Case 1: Automated Visual Field Testing (The Humphrey Field Analyzer). In standard static automated perimetry, the patient fixates inside an illuminated white bowl. The instrument maintains a continuous, uniform background adaptation luminance of exactly 31.5 asb (10 cd/m²). The maximum unattenuated test stimulus projected by the perimeter light source is calibrated to 10,000 asb (approximately 3,183 cd/m²). Clinically, stimulus brightness is modulated using neutral density filters measured in decibels (dB), where 10 dB represents a 10-fold (1.0 log unit) reduction in stimulus intensity. A stimulus of 0 dB corresponds to the full 10,000 asb intensity, 10 dB corresponds to 1,000 asb, 20 dB corresponds to 100 asb, 30 dB corresponds to 10 asb, and 40 dB corresponds to 1 asb. When a healthy fovea detects a stimulus at 35 dB, it is perceiving a target luminance of merely 3.16 apostilbs against the 31.5 asb background.

Case 2: Architectural Surface Reflection. Consider an office space lit with ceiling-mounted fluorescent or LED luminaires providing an illuminance of 500 lux on a work desk. If the desk surface is covered with matte white drafting paper featuring a diffuse reflection factor of ρ = 0.80 (80% reflectance), the resulting luminance of the paper calculated in apostilbs is simply 500 × 0.80 = 400 asb. In SI units, this corresponds to 400 / π ≈ 127.3 cd/m².

Case 3: Natural Environmental Luminance Levels. Under clear nighttime conditions under a full moon, outdoor ground illuminance is approximately 0.1 lux. A patch of clean, fresh snow (ρ ≈ 0.90) exhibits a luminance of approximately 0.09 asb (0.028 cd/m²), placing human vision firmly in the mesopic-to-scotopic transition zone. Conversely, an outdoor landscape under direct summer sunlight receiving 100,000 lux onto light-colored dry sand (ρ ≈ 0.40) produces a luminance of roughly 40,000 asb (≈ 12,732 cd/m²).

9. Measurement & Assessment

The empirical determination and calibration of luminance values expressed in apostilbs rely on high-precision optical instruments known as spot photometers, telephotometers, or luminance meters. Modern electronic luminance meters operate via calibrated silicon photodiodes paired with precision optical bandpass filters that meticulously correct the incoming spectral responsivity to mirror the CIE 1924 standard photopic observer curve V(λ).

To measure luminance in apostilbs directly using modern equipment:

  • The photometer measures the directional luminance in the standard SI unit of candelas per square meter (cd/m²).
  • The measured SI value is subsequently converted by multiplying the reading by π (≈ 3.14159) to yield the corresponding value in apostilbs.
  • In clinical settings, automated perimeters contain automated internal self-calibration photometers. Prior to initiating examination routines, an internal photodiode samples both the background bowl illumination and the target projection beam, compensating for projector lamp aging, dust accumulation on the bowl surface, and power fluctuations to ensure the adaptation background stays within a tight tolerance of 31.5 asb ± 10%.
  • For calibration verification against primary national metrology standards, laboratory physicists utilize diffuse reflectance standards made of pressed polytetrafluoroethylene (PTFE) or sintered barium sulfate (BaSO4, commercially known as Spectralon), which exhibit near-perfect Lambertian reflectance (ρ > 0.99) across the entire visible spectrum.

10. Applications & Practical Significance

Although largely replaced in commercial engineering, the apostilb retains critical significance across specialized branches of clinical healthcare and scientific history:

  • Clinical Glaucoma Management: Glaucoma induces progressive, irreversible apoptotic damage to retinal ganglion cells, leading to characteristic visual field scotomas. The progression of functional loss is universally tracked via automated perimeters calibrated in apostilbs and decibels. Treatment decisions—such as escalating ocular hypotensive medications or performing trabeculectomy—depend directly on threshold sensitivity values calibrated against the 31.5 asb standard.
  • Neuro-Ophthalmological Evaluation: Compressive lesions of the optic chiasm (such as pituitary adenomas causing bitemporal hemianopsia) and ischemic strokes affecting the retrochiasmal visual pathway are mapped using kinetic Goldmann perimetry. Target size and luminance gradations in Goldmann testing utilize Roman numerals (target size) and Arabic numbers/letters (target luminance in apostilb steps), enabling standardized cross-institutional comparisons over decades of patient records.
  • Visual Psychophysics and Spatial Vision Research: In research investigating dark adaptation curves (e.g., measuring rod-cone break times), contrast sensitivity, and temporal flicker fusion frequencies, legacy databases and contemporary studies frequently reference stimulus values in apostilbs to maintain parity with historical psychophysical literature.
  • Historical Ergonomics and German Industrial Standards: Anyone reviewing mid-century European architectural archives, factory illumination blueprints, and early human factors guidelines encounters workplace illumination requirements specified in apostilbs. Familiarity with the unit remains necessary for architectural historians and occupational health researchers evaluating past working conditions.

11. Research & Empirical Evidence

The empirical foundation underpinning the selection of 31.5 apostilbs as the global standard for perimetric background luminance rests on landmark studies in visual physiology conducted during the early and mid-twentieth century. Researchers including Selig Hecht, H. Richard Blackwell, and Hans Goldmann investigated how background luminance impacts the differential light threshold (ΔI).

Under scotopic conditions (low luminance, below 0.01 asb), the visual system is rod-dominated, spatial acuity is poor, and the contrast threshold is non-linear. As background luminance climbs into the photopic regime (above 10 asb), rod photoreceptors saturate, and cone-mediated vision governs the visual response. In his classic 1945 monographs, Goldmann demonstrated that at a background luminance of approximately 10 cd/m² (31.5 asb), the eye operates comfortably on the flat portion of the Weber contrast curve:

ΔL / L = constant

This stability is critical: at 31.5 asb, minor involuntary pupil fluctuations, modest room ambient light leaks, or mild variations in lamp output do not alter the patient's relative contrast threshold. Furthermore, this adaptation level maintains pupil size in an intermediate state, avoiding both extreme miosis (which introduces severe optical diffraction) and extreme mydriasis (which increases spherical optical aberrations).

Subsequent multi-center clinical trials validating automated perimetry algorithms, such as the Swedish Interactive Thresholding Algorithm (SITA) and the Humphrey Field Analyzer development trials led by Anders Heijl, Boel Bengtsson, and Vincent Michael Patella, preserved the 31.5 asb background and 10,000 asb maximum stimulus standards. By preserving this calibration baseline, massive longitudinal cohorts in the Early Manifest Glaucoma Trial (EMGT) and the Ocular Hypertension Treatment Study (OHTS) maintained rigorous statistical comparability over decades of patient follow-up.

12. Cultural & Cross-Cultural Considerations

The utilization and eventual obsolescence of the apostilb reflect historical geopolitical divisions in scientific metrology. During the mid-twentieth century, a distinct divide existed between continental European engineering and Anglo-American practice:

  • Continental Europe: Led by German, French, and Swiss scientific societies, continental Europe broadly embraced metric diffuse luminance units, establishing the apostilb and the blondel across industry, architecture, and clinical medicine. The standard German industrial illumination norm DIN 5035 entrenched the apostilb across thousands of technical workplaces.
  • The Anglosphere: British and American illumination engineers steadfastly preferred imperial units, relying predominantly on the foot-lambert (fL) and the candle per square inch. In the United States, military technical specifications and illumination guidelines published by the Illuminating Engineering Society of North America (IESNA) favored the foot-lambert until the SI system mandated candelas per square meter.
  • Ophthalmic Universalism: Because Haag-Streit (Switzerland) held a near-monopoly on high-end precision optical perimeters following World War II, the European standard of 31.5 asb bypassed nationalistic engineering barriers. American and British ophthalmologists adopted the Swiss standard universally, leading to an intriguing historical paradox: while American engineering rejected the apostilb, American clinical ophthalmology adopted and maintained it for clinical practice.

13. Criticisms, Debates & Limitations

Despite its mathematical elegance for diffuse Lambertian scenarios, the apostilb faced intense theoretical criticism throughout the latter half of the twentieth century, which ultimately motivated its formal retirement from modern physics and engineering.

The most substantial theoretical criticism centers on the conceptual confusion between luminance and luminous exitance. By embedding the factor of π into the unit to make its numerical value equal to exitance (lumens/m²), the apostilb blurred the fundamental dimensional and physical distinctions between isotropic spatial flux emission and directional intensity per projected area. In non-Lambertian environments—such as glossy surfaces, retroreflective street signs, polished metals, or digital display screens—the assumption of uniform hemispherical diffusion fails completely. Applying an exitance-derived unit like the apostilb to non-diffuse, specular surfaces creates significant mathematical errors.

A second major limitation involves unit proliferation and operational redundancy. Metrological organizations emphasized that maintaining multiple coexisting units for the same physical quantity (candela per square meter, nit, stilb, lambert, foot-lambert, apostilb, skot, and bril) generated unnecessary confusion in international trade, academic research, and instrumentation manufacturing. The adoption of the coherent SI unit, candela per square meter (cd/m²), rendered all such auxiliary units mathematically obsolete.

Finally, in modern clinical perimetry, the ongoing retention of the apostilb alongside logarithmic decibel (dB) scales has created widespread confusion among practicing clinicians. Because perimeters use inverted logarithmic attenuation scales where 0 dB represents maximum instrument brightness (e.g., 10,000 asb) and 30 dB represents a faint stimulus (10 asb), many ophthalmology trainees struggle to reconcile the inverse decibel scale with physical luminance measured in apostilbs or candelas per square meter.

14. Related Terms & Distinctions

To prevent confusion in optical and ophthalmic contexts, the apostilb must be distinguished from related photometric terms:

  • Apostilb vs. Candela per Square Meter (cd/m² or Nit): The candela per square meter is the coherent SI unit of luminance. It measures directional luminous intensity per unit projected area without assuming diffuse reflection. One apostilb equals 1/π cd/m² (≈ 0.318 cd/m²). The nit is an informal synonym for cd/m².
  • Apostilb vs. Stilb (sb): The stilb is the CGS unit of luminance, defined as 1 candela per square centimeter (10,000 cd/m²). The apostilb is equal to exactly (10−4/π) stilbs, making it roughly 31,416 times smaller than a stilb.
  • Apostilb vs. Lambert (L): The lambert is the CGS diffuse luminance unit, defined as 1/π cd/cm². One lambert equals exactly 10,000 apostilbs.
  • Apostilb vs. Foot-Lambert (fL): The foot-lambert is the imperial diffuse luminance unit, defined as 1/π cd/ft². One foot-lambert equals approximately 10.764 apostilbs.
  • Apostilb vs. Lux (lx): Lux is the SI unit of illuminance (incident luminous flux per area, lumens/m²). While an apostilb is numerically equivalent to a lux when viewing a perfectly reflecting Lambertian surface, lux describes incident light striking a surface, whereas apostilb describes light reflected or emitted from it.
  • Apostilb vs. Troland (Td): The troland is a physiological unit of retinal illuminance, equal to the surface luminance in cd/m² multiplied by pupil area in square millimeters. The troland incorporates the physiological aperture of the human eye, whereas the apostilb measures physical surface emission independent of the observer.

15. Summary / Key Takeaways

The apostilb is a historical non-SI photometric unit of luminance designed to simplify calculations for perfectly diffuse, Lambertian surfaces. Formally defined as (1/π) candela per square meter, one apostilb corresponds to the luminance of an ideal reflector emitting one lumen per square meter. Although the International System of Units officially deprecated the apostilb in favor of the candela per square meter (cd/m²) due to metrological standardization and the conceptual pitfalls of treating luminance as exitance, the unit remains globally relevant in clinical visual science. Through its foundational adoption in the Goldmann perimeter and modern automated perimetry, the apostilb endures as an essential reference in the diagnosis, staging, and longitudinal management of vision-threatening ophthalmic disorders.

References

  • Anderson, D. R., & Patella, V. M. (1999). Automated static perimetry (2nd ed.). Mosby.
  • Blondel, A. (1896). Sur la photometrie des grandeurs lumineuses continues et discontinues. Journal de Physique Théorique et Appliquée, 5(1), 389–398.
  • Bureau International des Poids et Mesures. (2019). The International System of Units (SI) (9th ed.). International Bureau of Weights and Measures.
  • Commission Internationale de l'Éclairage. (2020). International Lighting Vocabulary (CIE S 017/E:2020). CIE Central Bureau.
  • Goldmann, H. (1945). Grundlagen exakter Perimetrie. Ophthalmologica, 109(2-3), 57–70.
  • Heijl, A., Lindgren, G., & Olsson, J. (1987). A package for the statistical analysis of computerized fields. Documenta Ophthalmologica Proceedings Series, 49, 153–168.
  • Walsh, J. W. T. (1958). Photometry (3rd ed.). Constable & Company Ltd.

Cite This Article

Looti, M. (2026, November 6). Apostilb: Luminance Measurement in Vision Science. ASHA PSYCHOLOGICAL DATABASE. https://ashadocs.org/?psych_term=apostilb-luminance-measurement-vision-science
Looti, Mohammed. “Apostilb: Luminance Measurement in Vision Science.” ASHA PSYCHOLOGICAL DATABASE, 6 November 2026, https://ashadocs.org/?psych_term=apostilb-luminance-measurement-vision-science.
Looti, Mohammed. “Apostilb: Luminance Measurement in Vision Science.” ASHA PSYCHOLOGICAL DATABASE. November 6, 2026. https://ashadocs.org/?psych_term=apostilb-luminance-measurement-vision-science.